MISC

2014年

Development of a Microcalorimeter for Measurement of L X-rays Emitted from Transuranium Elements

2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC)
  • K. Nakamura
  • ,
  • M. Maeda
  • ,
  • K. Maehata
  • ,
  • N. Iyomoto
  • ,
  • K. Ishibashi
  • ,
  • K. Takasaki
  • ,
  • K. Mitsuda

記述言語
英語
掲載種別
DOI
10.1109/NSSMIC.2014.7431099
出版者・発行元
IEEE

We fabricated a transition-edge sensor (TES) microcalorimeter for spectroscopic measurements of L X-rays emitted by transuranium (TRU) elements. A Pu L X-ray energy spectrum was obtained by irradiation with L X-rays emitted by a 244 Cm source to the TES microcalorimeter. The energy resolution was evaluated to be 64 eV at the full width at half maximum (FWHM) for L X-rays at 14.28 keV. The obtained L X-ray spectrum indicated that the TES microcalorimeter made it possible to distinguish between the L X-ray lines of TRU elements.