Predictive factors of truncation artifacts in the arterial phase of Gd-EOB-DTPA-enhanced MRI: a nationwide multicenter study.
Japanese journal of radiology
- ,
- ,
- ,
- ,
- ,
- ,
- ,
- ,
- 巻
- 39
- 号
- 2
- 開始ページ
- 165
- 終了ページ
- 177
- 記述言語
- 英語
- 掲載種別
- 研究論文(学術雑誌)
- DOI
- 10.1007/s11604-020-01052-x
PURPOSE: To identify predictive factors for truncation artifacts (TAs) in the arterial phase of Gd-EOB-DTPA-enhanced MRI in a multicenter study in Japan. MATERIALS AND METHODS: Data on patient factors (age, sex, weight, presence of viral hepatitis, and other conditions) and imaging parameters (e.g., triggering, voxel size, matrix, k-space ordering, acquisition time, reduction factor, flip angle, fat suppression, field strength, injection rate, and saline volume) were obtained. Univariate and multivariate analyses were performed to investigate the correlation of these parameters. RESULTS: We evaluated 1444 patients from 43 institutions who were scanned using GE, Siemens, Philips, or Toshiba MRI equipment (501, 354, 349, and 240 patients, respectively). The total incidence of TAs was 12.5% (17.2, 3.6, 15.7, and 12.1%, respectively). The matrix [odds ratio (OR) 0.13], flip angle (OR 5.77), use of fat suppression (OR 0.106), and field strength (OR 0.092) used in the Philips equipment significantly increased the incidence of TAs in MRI examination. CONCLUSIONS: The incidence of TAs in the arterial phase is influenced by several patient factors and imaging parameters. Especially, Siemens and Toshiba equipment had a significantly lower frequency of TAs. This indicates that such vendor-specific technology used in the dynamic sequence may have a TA-resistant effect.
- リンク情報
- ID情報
-
- DOI : 10.1007/s11604-020-01052-x
- PubMed ID : 33025339