Analytical validation on carbon and oxygen isotopic measurement of small carbonate samples by using IsoPrime100 mass spectrometer
CARBONATES AND EVAPORITES
- ,
- 巻
- 32
- 号
- 1
- 開始ページ
- 117
- 終了ページ
- 122
- 記述言語
- 英語
- 掲載種別
- 研究論文(学術雑誌)
- DOI
- 10.1007/s13146-015-0279-9
- 出版者・発行元
- SPRINGER
This paper focused on analytical validation such as the amount needed, time drift, repeatability, memory effect, analytical precision and accuracy on carbon/oxygen isotopic measurement of small carbonate samples (10-80 A mu g) using an IsoPrime100 mass spectrometer. 10 A mu g of carbonate samples yielded mass 44-intensity of 2 nA. The minimum amount needed was 10 mu g and the repeatability was around 0.1 aEuro degrees (1 s.d.). Through 24-h measurement there was no temporal variation of delta C-13, but that of delta O-18 showed variation with decreasing the amount of CO2 reference gas, suggesting that delta O-18 values need to conduct time drift correction. Inter-sample memory effect was not observed, indicating that IsoPrime100 mass spectrometer with CO2 purification system is adequately evacuated after each analysis by means of pumping system. Moreover, IAEA-CO-1 and JAMSTEC-WS values calibrated by international reference NBS19 agree within error with the reference values, demonstrating the accuracy of our analytical procedure on carbon/oxygen isotopic measurement of carbonate samples. This paper offers insights into analytical validation procedure when setting up a mass spectrometer on isotopic analyses of small carbonate samples.
- リンク情報
- ID情報
-
- DOI : 10.1007/s13146-015-0279-9
- ISSN : 0891-2556
- eISSN : 1878-5212
- ORCIDのPut Code : 117901625
- Web of Science ID : WOS:000396032800010