論文

査読有り
2015年12月

Minimum 10-Year Wear Analysis of Highly Cross-Linked Polyethylene in Cementless Total Hip Arthroplasty

JOURNAL OF ARTHROPLASTY
  • Kazutaka So
  • ,
  • Koji Goto
  • ,
  • Yutaka Kuroda
  • ,
  • Shuichi Matsuda

30
12
開始ページ
2224
終了ページ
2226
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1016/j.arth.2015.06.053
出版者・発行元
CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS

Fifty-four patients (64 hips) underwent cementless total hip arthroplasty between 2000 and 2003 with a 22-mm zirconia ceramic bearing on highly cross-linked polyethylene, and were evaluated with a mean 11.9-year postoperative follow-up (range, 10-14 years). Linear wear was measured on the anteroposterior radiograph of the hip. No evidence of osteolysis and loosening was found on the final radiograph in any of the cases, and the steady-state linear wear rate was 0.017 +/- 0.018 mm/year. No significant correlation was found between the linear wear rate and age, body weight, cup inclination angle, or polyethylene thickness. Highly cross-linked polyethylene showed excellent wear resistance for >10 years when used in combination with 22-mm zirconia heads. (C) 2015 Elsevier Inc. All rights reserved.

リンク情報
DOI
https://doi.org/10.1016/j.arth.2015.06.053
PubMed
https://www.ncbi.nlm.nih.gov/pubmed/26231076
Web of Science
https://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=JSTA_CEL&SrcApp=J_Gate_JST&DestLinkType=FullRecord&KeyUT=WOS:000366161300035&DestApp=WOS_CPL
Scopus
https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=84952873583&origin=inward
Scopus Citedby
https://www.scopus.com/inward/citedby.uri?partnerID=HzOxMe3b&scp=84952873583&origin=inward
ID情報
  • DOI : 10.1016/j.arth.2015.06.053
  • ISSN : 0883-5403
  • eISSN : 1532-8406
  • PubMed ID : 26231076
  • SCOPUS ID : 84952873583
  • Web of Science ID : WOS:000366161300035

エクスポート
BibTeX RIS