Evaluation of apparent diffusion coefficient of two-dimensional BLADE turbo gradient- and spin-echo diffusion-weighted imaging with a breast phantom.
Radiological physics and technology
- ,
- ,
- ,
- ,
- ,
- 巻
- 16
- 号
- 1
- 開始ページ
- 118
- 終了ページ
- 126
- 記述言語
- 英語
- 掲載種別
- 研究論文(学術雑誌)
- DOI
- 10.1007/s12194-022-00694-y
This study aimed to evaluate the reliability of apparent diffusion coefficient (ADC) values generated with two-dimensional turbo gradient- and spin-echo with BLADE trajectory diffusion-weighted imaging (TGSE-BLADE-DWI) sequence using a breast diffusion phantom. TGSE-BLADE-DWI and single-shot spin-echo echo-planar imaging (SS-EPI-DWI) were performed using a 3.0 T magnetic resonance imaging scanner. Concordance rates of ADC values and the signal-to-noise ratio (SNR) were compared between TGSE-BLADE-DWI and SS-EPI-DWI. TGSE-BLADE-DWI provided a higher concordance rate for ADC values than SS-EPI-DWI when b-values > 2000s/mm2 and a slice thickness of 1 mm were used. TGSE-BLADE-DWI showed less image distortion than SS-EPI-DWI. The SNR of TGSE-BLADE-DWI was higher than that of SS-EPI-DWI, except at a number of excitations of 7 and a slice thickness of 1 mm. In conclusion, TGSE-BLADE-DWI can offer a better SNR, less distortion, and more reliable ADC measurements than SS-EPI-DWI in a breast phantom.
- リンク情報
- ID情報
-
- DOI : 10.1007/s12194-022-00694-y
- PubMed ID : 36596917