論文

査読有り 国際誌
2019年12月6日

Size-specific dose estimates for various weighting factors of CTDI equation.

Australasian physical & engineering sciences in medicine
  • Tomonobu Haba
  • ,
  • Masanao Kobayashi
  • ,
  • Shuji Koyama

記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1007/s13246-019-00830-w

Size-specific dose estimate (SSDE) was proposed by the American Association of Physicists in Medicine (AAPM) Task Group 204 to consider the effect of patient size in the x-ray CT dose estimation. Size correction factors to calculate SSDE were derived based on the conventional weighted CT dose index (CTDIw) equation. This study aims to investigate the influence of Bakalyar's and the authors' own CTDIw equations on the size correction factors described by the AAPM Task Group 204, using Monte Carlo simulations. The simulations were performed by modeling four types of x-ray CT scanner designs, to compute the dose values in water for cylindrical phantoms with 8-40 cm diameters. CTDI100 method and the AAPM Task Group 111's proposed method were employed as the CT dosimetry models. Size correction factors were obtained for the computed dose values of various phantom diameters for the conventional, Bakalyar's, and the authors' weighting factors. Maximum difference between the size correction factors for the Bakalyar's weighting factor and those of the AAPM Task Group 204 was 27% for a phantom diameter of 11.2 cm. On the other hand, the size correction factors calculated for the authors' weighting factor were in good agreement with those from the AAPM Task Group 204 report with a maximum difference of 17%. The results indicate that the SSDE values obtained with the authors' weighting factor can be evaluated by using the size correction factors reported by the AAPM Task Group 204, which is currently accepted as a standard.

リンク情報
DOI
https://doi.org/10.1007/s13246-019-00830-w
PubMed
https://www.ncbi.nlm.nih.gov/pubmed/31811559
ID情報
  • DOI : 10.1007/s13246-019-00830-w
  • PubMed ID : 31811559

エクスポート
BibTeX RIS