Spatial resolution in plantar pressure measurement revisited

JOURNAL OF BIOMECHANICS
  • Todd C. Pataky

45
12
開始ページ
2116
終了ページ
2124
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1016/j.jbiomech.2012.05.038
出版者・発行元
ELSEVIER SCI LTD

Plantar pressures are typically measured using sensors of finite area, so the accuracy with which one can measure true maximum pressure is dependent on sensor size. Measurement accuracy has been modeled previously for one patient's metatarsals (Lord, 1997), but has not been modeled either for general subjects or for other parts of the foot. The purposes of this study were (i) to determine whether Lord's (1997) model is also valid for heel and hallux pressures, and (ii) to examine how sensor size relates to measurement accuracy in the context of four factors common to many measurement settings: pressure pulse size, foot positioning, pressure change quantification, and gross pressure redistribution. Lord's (1997) model was first generalized and was then validated using 10 healthy walking subjects, with relatively low RMSE values on the order of 20 kPa. Next, postural data were used to show that gross pressure redistributions can be accurately quantified (p < 0.002). even with rather gross sensor sizes of 30 mm. Finally, numerical analyses revealed that the relation between sensor size and measurement accuracy is highly complex, with deep dependency on the measurement context. In particular, the critical sensor widths required to achieve 90% accuracy ranged from 1.7 mm to 17.4 mm amongst the presently investigated scenarios. Since measurement accuracy varies so extensively with so many factors, the current results cannot yield specific recommendations regarding spatial resolution. It is concluded simply that no particular spatial resolution can yield a constant measurement accuracy across common plantar pressure measurement tasks. (C) 2012 Elsevier Ltd. All rights reserved.

リンク情報
DOI
https://doi.org/10.1016/j.jbiomech.2012.05.038
Web of Science
https://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=JSTA_CEL&SrcApp=J_Gate_JST&DestLinkType=FullRecord&KeyUT=WOS:000307905800009&DestApp=WOS_CPL
ID情報
  • DOI : 10.1016/j.jbiomech.2012.05.038
  • ISSN : 0021-9290
  • Web of Science ID : WOS:000307905800009

エクスポート
BibTeX RIS