2018年7月
Consideration of Calculation Process Assuming Heart Rate Variability Analysis Using Wearable ECG Devices
The 40th Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC'18)
- ,
- ,
- ,
- ,
- 巻
- 2018
- 号
- 開始ページ
- 5693
- 終了ページ
- 5696
- 記述言語
- 英語
- 掲載種別
- 研究論文(国際会議プロシーディングス)
- DOI
- 10.1109/EMBC.2018.8513449
- 出版者・発行元
- IEEE
Since many shirt-type wearable electrocardiogram (ECG) devices employ shirt-embedded dry electrodes that are easily affected by the daily activities of users, measured ECG often includes noise or artifacts. Experiments confirmed that a desirable HRV calculation process assuming these wearable ECG devices is a combination of accurate R wave detection, RRI outlier exclusion and missing RRI complement methods.
- リンク情報
- ID情報
-
- DOI : 10.1109/EMBC.2018.8513449
- PubMed ID : 30441628