2017年2月
Quantitative assessment technique of HyperEye medical system angiography for coronary artery bypass grafting
SURGERY TODAY
- ,
- ,
- ,
- ,
- ,
- ,
- ,
- 巻
- 47
- 号
- 2
- 開始ページ
- 210
- 終了ページ
- 217
- 記述言語
- 英語
- 掲載種別
- 研究論文(学術雑誌)
- DOI
- 10.1007/s00595-016-1369-6
- 出版者・発行元
- SPRINGER
The HyperEye Medical System (HEMS) uses indocyanine green (ICG) to visualize blood vessels in coronary artery bypass grafting (CABG). We performed quantitative HEMS assessment to detect grafts at risk of occlusion.
We assessed the HEMS angiograms of 177 grafts from 69 patients who underwent CABG and compared the results with those of fluoroscopic coronary angiography, by measuring the increasing rate of ICG intensity, average acceleration value, and time to peak luminance intensity.
Grafts in the patent and failed groups showed significant differences in their increasing rate of intensity and average acceleration value. The average accelerations value of ICG intensity of internal thoracic artery (ITA) and saphenous vein (SV) grafts were 112.3 and 144.9 intensity/s(2) in the patent group, and 71.0 and 91.8 intensity/s(2) in the failed group. The time to peak luminance intensity was 1.7 and 1.4 s in the patent group and 2.3 and 1.9 s in the failed group; these values were not significantly different.
Significant reductions in the ICG intensity rate and average acceleration value can occur in failed grafts. Therefore, quantifiable changes in ICG intensity may help detect minute changes in blood flow.
We assessed the HEMS angiograms of 177 grafts from 69 patients who underwent CABG and compared the results with those of fluoroscopic coronary angiography, by measuring the increasing rate of ICG intensity, average acceleration value, and time to peak luminance intensity.
Grafts in the patent and failed groups showed significant differences in their increasing rate of intensity and average acceleration value. The average accelerations value of ICG intensity of internal thoracic artery (ITA) and saphenous vein (SV) grafts were 112.3 and 144.9 intensity/s(2) in the patent group, and 71.0 and 91.8 intensity/s(2) in the failed group. The time to peak luminance intensity was 1.7 and 1.4 s in the patent group and 2.3 and 1.9 s in the failed group; these values were not significantly different.
Significant reductions in the ICG intensity rate and average acceleration value can occur in failed grafts. Therefore, quantifiable changes in ICG intensity may help detect minute changes in blood flow.