2011年2月
Note: Accuracy of velocity correction for impact of a laser-accelerated miniature flyer with lithium fluoride shock-compressed along the [100] axis
REVIEW OF SCIENTIFIC INSTRUMENTS
- ,
- ,
- ,
- 巻
- 82
- 号
- 2
- 開始ページ
- 026112
- 終了ページ
- 記述言語
- 英語
- 掲載種別
- 研究論文(学術雑誌)
- DOI
- 10.1063/1.3553293
- 出版者・発行元
- AMER INST PHYSICS
We performed miniature flyer impact experiments to investigate the relationship between the apparent (u(a)) and actual (u(A)) particle velocities measured by a velocity interferometer in single-crystal lithium fluoride (LiF) that was shock-compressed along the [100] axis. The miniature flyer was accelerated to velocities in the range 652.5-1937.6 m/s by a tabletop pulsed laser. An empirical relationship of u(a) = (1.2749 +/- 0.0102)u(A) was obtained. The obtained relationship agreed well with the results of a previous study within the experimental errors and its uncertainty was less than +/- 1%. This result indicates that the present experimental technique is effective for measuring the relationship between u(a) and u(A) of shocked transparent materials with a comparable accuracy to conventional methods. (C) 2011 American Institute of Physics. [doi:10.1063/1.3553293]
- リンク情報
- ID情報
-
- DOI : 10.1063/1.3553293
- ISSN : 0034-6748
- eISSN : 1089-7623
- PubMed ID : 21361651
- Web of Science ID : WOS:000287813400085