Computed tomography reconstruction from two transmission measurements for iodine-marked cancer detection

JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY
  • Ikuo Kanno
  • ,
  • Kazunari Shima
  • ,
  • Hironobu Shimazaki
  • ,
  • Yoshiki Yamashita
  • ,
  • Kenichi Watanabe
  • ,
  • Masahiko Ohtaka
  • ,
  • Makoto Hashimoto
  • ,
  • Kuniaki Ara
  • ,
  • Hideaki Onabe

50
10
開始ページ
1020
終了ページ
1033
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1080/00223131.2013.830077
出版者・発行元
TAYLOR & FRANCIS LTD

The authors invented the transXend detector, which measures X-rays as electric currents, and then gives the energy distribution of the X-rays after an unfolding process. In a previous paper, it was shown that the material thickness distributions can be estimated with the transXend detector by using reference points plotted from the electric current ratios, such as the I-2/I-1 - I-3/I-1 graph, where I-i denotes the electric current measured by the i-th segment of the transXend detector. In this paper, the tomographic images of iodine, aluminum, and the acrylic those surround the other two materials are reconstructed from their material thickness distributions, which are estimated from two X-ray incidence directions. The X-ray event ratios are also used to estimate the material thickness distributions.