論文

査読有り
2019年6月

Nanoneedle-Electrode Array Packaged with Amplifiers for Recording Biological-Signals with A High Voltage Gain

2019 20th International Conference on Solid-State Sensors, Actuators and Microsystems and Eurosensors XXXIII, TRANSDUCERS 2019 and EUROSENSORS XXXIII
  • Shuhei Tsuruhara
  • ,
  • Yoshihiro Kubota
  • ,
  • Hiroshi Kubo
  • ,
  • Hirohito Sawahata
  • ,
  • Shota Yamagiwa
  • ,
  • Shinnosuke Idogawa
  • ,
  • Takeshi Kawano

開始ページ
1784
終了ページ
1787
記述言語
掲載種別
研究論文(国際会議プロシーディングス)
DOI
10.1109/TRANSDUCERS.2019.8808366

We have proposed nanoscale-tipped microneedle-electrodes by vapor-liquid-solid growth of silicon microwire and the subsequent sharpening process. However, due to the small area of the recording site (< 1 μm in diameter), the needle exhibits high electrical impedance characteristics, causing amplitude attenuation of recorded signals at > 100 Hz. To overcome the device issue, here we designed and fabricated a buffer amplifier using a 1×1 mm silicon-block. Assembling the buffer amplifier improves the gain of the nanoneedle device, as confirmed in the test signal recording. These results suggest that the nanoneedle device with buffer amplifier enables recording of high signal amplitude of neuron and cell. 2