論文

査読有り
2018年11月

Sterilization of small vial using electron cyclotron resonance plasma

VACUUM
  • Tatsuya Nishikawa
  • ,
  • Norihito Abe
  • ,
  • Akira Yonesu
  • ,
  • Nobuya Hayashi

157
開始ページ
100
終了ページ
104
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1016/j.vacuum.2018.06.069
出版者・発行元
PERGAMON-ELSEVIER SCIENCE LTD

The sterilization of a small vial using electron cyclotron resonance (ECR) plasma was investigated. The ECR plasma was generated throughout the inside of the small vial, which was sealed with nonwoven fabric, by using magnetic configuration in which the ECR point was located inside the small vial. Moreover, it was demonstrated that the temperature of the small vial was finely controlled by the pulse width and pulse frequency of the pulsed microwave power. A biological indicator (BI), Geobacillus Stearothermophilus, containing 1 x 10(5) spores present inside the small vial was successfully inactivated in a total treatment time of about 14 min while maintaining treatment temperature of less than 65 degrees C.

リンク情報
DOI
https://doi.org/10.1016/j.vacuum.2018.06.069
Web of Science
https://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=JSTA_CEL&SrcApp=J_Gate_JST&DestLinkType=FullRecord&KeyUT=WOS:000449569600016&DestApp=WOS_CPL
ID情報
  • DOI : 10.1016/j.vacuum.2018.06.069
  • ISSN : 0042-207X
  • Web of Science ID : WOS:000449569600016

エクスポート
BibTeX RIS