Photoneutron-induced damage reduction for cardiac implantable electronic devices using neutron-shielding sheets in high-energy X-ray radiotherapy: A phantom study

Physica Medica
  • Ryo Kakino
  • ,
  • Mitsuhiro Nakamura
  • ,
  • Naonori Hu
  • ,
  • Hiraku Iramina
  • ,
  • Hiroki Tanaka
  • ,
  • Yoshinori Sakurai
  • ,
  • Takashi Mizowaki

89
開始ページ
151
終了ページ
159
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1016/j.ejmp.2021.07.036
出版者・発行元
Elsevier BV

PURPOSE: To evaluate damage reduction in cardiac implantable electronic devices (CIEDs) caused by photoneutrons in high-energy X-ray radiotherapy using a neutron-shielding sheet (NSS). METHODS: The NSS consists of a bolus with a thickness of 1 or 2 cm (Bls1 or Bls2) as a moderator and several absorbers (20%, 50%, or 80% B4C silicone sheet [B4C20, B4C50, or B4C80] or a 40% LiF silicone sheet [LiF40]). First, a linear accelerator (LINAC) with a water-equivalent phantom was modeled in the simulation and measured experimentally. Several NSSs were placed on the phantom, a Eu:LiCaAlF6 scintillator was placed between the phantom and the NSS, and X-rays were irradiated. The relative counts (Cr = counts when placing the NSS or Bls2) were compared between the experiment and simulation. Second, CIED damage was evaluated in the simulation. The relative damage (Dr = damage when placing or not placing the NSS) was compared among all the NSSs. In addition, the γ-ray and leaking X-ray dose from B4C was measured using a dosimetric film. After determining the optimal NSS combination, Dr value analysis was performed by changing the length of one side and the thickness. RESULTS: The Cr values of the simulation and experiment agreed within a 30% percentage difference, except for Bare or LiF40-only. The Dr value was reduced by 43% when Bls2 + B4C80 was applied. The photon dose was less than 5 cGy/1500 MU. The Dr values were smaller for the smaller lengths of one side of B4C80 and decreased as the M-layer thickness increased. CONCLUSIONS: The CIED damage induced by photoneutrons generated by a LINAC was effectively reduced by applying the optimal NSS.

リンク情報
DOI
https://doi.org/10.1016/j.ejmp.2021.07.036
PubMed
https://www.ncbi.nlm.nih.gov/pubmed/34371340
ID情報
  • DOI : 10.1016/j.ejmp.2021.07.036
  • ISSN : 1120-1797
  • PubMed ID : 34371340

エクスポート
BibTeX RIS