Misc.

1989

Helium-filled proportional counter for low-temperature operation (1.75-4.2 K) and its application to cryogenic resonance-electron Mössbauer spectroscopy II

Review of Scientific Instruments
  • Yasuhito Isozumi
  • ,
  • Shin Ito
  • ,
  • Tatuo Fujii
  • ,
  • Rintaro Katano

Volume
60
Number
10
First page
3262
Last page
3264
Language
English
Publishing type
DOI
10.1063/1.1140562

The operation of the helium-filled proportional counter (HFPC) has been investigated at low temperatures below liquid-helium temperature, 4.2 K. It has been found that the gas gain as a function of the anode voltage is not appreciably changed by the temperature of the filling gas, but continuous electric discharges are induced at lower anode voltages as the gas temperature is decreased. The present counter technique has been applied to resonance-electron Mössbauer spectroscopy in the temperature region below 4.2 K.

Link information
DOI
https://doi.org/10.1063/1.1140562
Scopus
https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=0343078297&origin=inward
Scopus Citedby
https://www.scopus.com/inward/citedby.uri?partnerID=HzOxMe3b&scp=0343078297&origin=inward
ID information
  • DOI : 10.1063/1.1140562
  • ISSN : 0034-6748
  • SCOPUS ID : 0343078297

Export
BibTeX RIS