A novel cup with a pressure-adjusting mechanism for high-temperature water vapor transmission rate measurements

POLYMER TESTING
  • Shinya Iizuka
  • ,
  • Kazuhide Murata
  • ,
  • Masahiro Sekine
  • ,
  • Chiaki Sato

50
開始ページ
73
終了ページ
78
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1016/j.polymertesting.2015.11.018
出版者・発行元
ELSEVIER SCI LTD

Water vapor transmission rate (WVTR) measurements with the conventional cup method do not yield accurate values at high temperatures because the film specimens deform and are damaged owing to air expansion in the cup. A new cup with a pressure-adjusting mechanism allows measurements at 85 degrees C and prevents specimen deformation and damage. WVTRs of polypropylene (PP) and polyethylene terephthalate (PET) measured with the new cup method are the same as those measured with the conventional cup method at 40 degrees C, and gas chromatographic detection method at 60 degrees C and 85 degrees C. Arrhenius plots of the water vapor permeability coefficient of PP, polyethylene naphthalate (PEN) and polyimide (PI) with the new cup method show a linear relationship in the range 25-85 degrees C. In the same range, Arrhenius plots of PET, polybutylene terephthalate (PBT) and polylactic acid (PIA) have bending points corresponding to the glass transition temperatures of the materials. (C) 2015 Elsevier Ltd. All rights reserved.

リンク情報
DOI
https://doi.org/10.1016/j.polymertesting.2015.11.018
Web of Science
https://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=JSTA_CEL&SrcApp=J_Gate_JST&DestLinkType=FullRecord&KeyUT=WOS:000372690400011&DestApp=WOS_CPL
URL
http://orcid.org/0000-0001-8784-5353
ID情報
  • DOI : 10.1016/j.polymertesting.2015.11.018
  • ISSN : 0142-9418
  • eISSN : 1873-2348
  • ORCIDのPut Code : 21141175
  • Web of Science ID : WOS:000372690400011

エクスポート
BibTeX RIS