論文

査読有り
2004年12月

Design of a compact 4 pi Ge spectrometer for the measurement of neutron capture cross sections of minor actinides

JOURNAL OF NUCLEAR SCIENCE AND TECHNOLOGY
  • M Sugawara
  • J Goto
  • M Oshima
  • H Harada
  • K Furutaka
  • M Igashira
  • T Ohsaki
  • M Mizumoto
  • A Osa
  • M Koizumi
  • Y Toh
  • Y Nagai
  • K Kawade
  • 全て表示

41
12
開始ページ
1129
終了ページ
1137
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.3327/jnst.41.1129
出版者・発行元
ATOMIC ENERGY SOC JAPAN

A compact 4pi Ge spectrometer was designed for the measurement of neutron capture cross sections of minor actinides. It consists of two EUROBALL clusters and four clover detectors with BGO shields surrounding a sample as close as possible so as to absorb a few gamma-rays of high energy emitted after the neutron capture reaction with reasonable efficiency. This feature opens up the possibility to discriminate the desired nuclide from other impurities based on the measurement of excitation energy and determine the cross section accurately.: Overall performances such as total detection efficiencies for 3 cascade gamma-rays as well as monochromatic gamma-rays were simulated using Geant4. It turned out we could achieve the peak detection efficiency as large as 0.9% for 3 cascade gamma-rays, which is almost an order of magnitude larger than the value of the Gammasphere.

リンク情報
DOI
https://doi.org/10.3327/jnst.41.1129
Web of Science
https://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=JSTA_CEL&SrcApp=J_Gate_JST&DestLinkType=FullRecord&KeyUT=WOS:000227001800001&DestApp=WOS_CPL
ID情報
  • DOI : 10.3327/jnst.41.1129
  • ISSN : 0022-3131
  • Web of Science ID : WOS:000227001800001

エクスポート
BibTeX RIS