Spatiotemporal treadmill gait measurements using a laser range scanner: feasibility study of the healthy young adults
PHYSIOLOGICAL MEASUREMENT
- ,
- ,
- ,
- ,
- ,
- ,
- ,
- ,
- 巻
- 38
- 号
- 4
- 開始ページ
- N81
- 終了ページ
- N92
- 記述言語
- 英語
- 掲載種別
- 研究論文(学術雑誌)
- DOI
- 10.1088/1361-6579/aa63d1
- 出版者・発行元
- IOP PUBLISHING LTD
Objective: Spatio-temporal parameters are typically used for gait analysis. Although these parameters are measured by sophisticated systems such as 3D motion capture system or optoelectronic bars, these systems cannot be deployed easily because of their high costs, large space requirements and elaborate set-up. The purpose of this study is to develope a system for measuring spatiotemporal gait parameters using a laser range scanner during treadmill gait. Approach: To calculate accurate spatiotemporal parameters, the differences between the laser range scanner measured values and the reference values obtained from a 3D motion capture system were investigated in thirty subjects. From measurements in time and position at foot contact/off, adjustments to compensate for the differences in time and position were derived. Then, to determine the validity of the proposed system, values from the proposed system and the reference system were compared in four additional subjects. Main results: The results indicate that the data from the laser range scanner demonstrate certain differences in time and position compared with reference values. However, when compensation values were introduced, each spatiotemporal parameter correlated well with the reference values.