2021年7月1日
Effectiveness evaluation of extrapolation to frequency response of hydrophone sensitivity for measuring instantaneous acoustic pressure of diagnostic ultrasound
Japanese Journal of Applied Physics
- ,
- 巻
- 60
- 号
- SD
- 開始ページ
- SDDE14_1
- 終了ページ
- SDDE14_7
- 記述言語
- 英語
- 掲載種別
- 研究論文(学術雑誌)
- DOI
- 10.35848/1347-4065/abf4a5
- 出版者・発行元
- IOP Publishing
Precise measurement techniques of instantaneous acoustic pressure using the frequency response of hydrophone sensitivity have been investigated to evaluate the safety of diagnostic ultrasound. This technique requires the frequency range of sensitivities broader than bandwidth of the ultrasound. But available frequency range of sensitivities on a calibration certificate is limited. Extrapolation to the frequency range of certificated sensitivities might be an effective approach; however, the effectiveness of such extrapolation has not been sufficiently studied. In this paper, we propose a quantitative method for evaluating the effectiveness of such extrapolation, which is characterized by showing the narrowest frequency range of certificated hydrophone sensitivity necessary for precise measurement of instantaneous acoustic pressure in diagnostic ultrasound. The evaluation results showed that the extrapolation was generally effective for precise measurement using the certificated hydrophone sensitivity with a narrower frequency range than the bandwidth of the measured ultrasound.
- ID情報
-
- DOI : 10.35848/1347-4065/abf4a5
- ISSN : 0021-4922
- eISSN : 1347-4065
- ORCIDのPut Code : 91645885