2014年
Evaluation of a near-infrared-type contrast medium extravasation detection system using a swine model
Journal of Computer Assisted Tomography
- ,
- ,
- ,
- ,
- ,
- ,
- ,
- 巻
- 38
- 号
- 2
- 開始ページ
- 285
- 終了ページ
- 292
- 記述言語
- 英語
- 掲載種別
- 研究論文(学術雑誌)
- DOI
- 10.1097/RCT.0000000000000013
- 出版者・発行元
- Lippincott Williams and Wilkins
PURPOSE: To refine the development and evaluate the near-infrared (NIR) extravasation detection system and its ability to detect extravasation during a contrast-enhanced computed tomography (CT) examination. MATERIALS AND METHODS: The NIR extravasation detection system projects the NIR light through the surface of the human skin then, using its sensory system, will monitor the changes in the amount of NIR that reflected, which varies based on absorption properties.Seven female pigs were used to evaluate the contrast media extravasation detection system, using a 20-gauge intravenous catheter, when injected at a rate of 1 mL/s into 4 different locations just under the skin in the thigh section. Using 3-dimensional CT images, we evaluated the extravasations between time and volume, depth and volume, and finally depth and time to detect. RESULTS: We confirmed that the NIR light, 950-nm wavelength, used by the extravasation detection system is well absorbed by contrast media, making changes easy to detect. The average time to detect an extravasation was 2.05 seconds at a depth of 2.0 mm below the skin with a volume of 1.3 mL, 2.57 seconds at a depth between 2.1 and 5 mm below the skin and a volume of 3.47 mL, 10.5 seconds for depths greater than 5.1 mm and a volume of 11.1 mL. The detection accuracy was significantly deteriorated when the depth exceeded 5.0 mm (Tukey-Kramer, P <
0.05) CONCLUSIONS: The extravasation system detection system that is using NIR has a high level of detection sensitivity. The sensitivity enables the system to detect extravasation at depths less than 2 mm with a volume of 1.5 mL and at depths less than 5 mm with a volume of 3.5 mL. The extravasation detection system could be suitable for use during examinations. © 2014 by Lippincott Williams &
Wilkins.
0.05) CONCLUSIONS: The extravasation system detection system that is using NIR has a high level of detection sensitivity. The sensitivity enables the system to detect extravasation at depths less than 2 mm with a volume of 1.5 mL and at depths less than 5 mm with a volume of 3.5 mL. The extravasation detection system could be suitable for use during examinations. © 2014 by Lippincott Williams &
Wilkins.
- リンク情報
- ID情報
-
- DOI : 10.1097/RCT.0000000000000013
- ISSN : 1532-3145
- ISSN : 0363-8715
- PubMed ID : 24448506
- SCOPUS ID : 84897051152