論文

2001年3月

新素材カセッテの表面材質の画質に与える影響について

日本放射線技術学会雑誌
  • 岡本孝英
  • ,
  • 丸石博文
  • ,
  • 小川敬寿
  • ,
  • 森剛
  • ,
  • 山崎哲史
  • ,
  • 前島秀幸
  • ,
  • 吾子俊敬
  • ,
  • 鎌江春憲
  • ,
  • 伊知地宏志

57
3
開始ページ
337
終了ページ
343
記述言語
日本語
掲載種別
研究論文(学術雑誌)
DOI
10.6009/jjrt.KJ00001357527
出版者・発行元
公益社団法人 日本放射線技術学会

This report describes a comparison of cassette front material in terms of its effect on radiographic imaging. We compared aluminum, carbon fiber, and a new material (Hylite) in terms of their absorption, scatter rate (lead-disc method), MTF, and subjective appearance, using an extremity phantom. For scatter rate and MTF, we also carried out analysis using Monte Carlo simulation. Results indicated that the cassette using the new material was intermediate between the aluminum panel and CFRP panel in terms of weight, absorption rate, and scatter rate. We also found that the new panel material showed a visual difference from the aluminum panel when a lower tube voltage and thinner object were assigned.

リンク情報
DOI
https://doi.org/10.6009/jjrt.KJ00001357527
J-GLOBAL
https://jglobal.jst.go.jp/detail?JGLOBAL_ID=200902151839281242
CiNii Articles
http://ci.nii.ac.jp/naid/110003439435
CiNii Books
http://ci.nii.ac.jp/ncid/AN00197784
ID情報
  • DOI : 10.6009/jjrt.KJ00001357527
  • ISSN : 0369-4305
  • J-Global ID : 200902151839281242
  • CiNii Articles ID : 110003439435
  • CiNii Books ID : AN00197784

エクスポート
BibTeX RIS