1993年11月1日
Micro Laser Ablation-Inductively Coupled Plasma Mass Spectrometry. 1. Instrumentation and Performance of Micro Laser Ablation System
Analytical Chemistry
- ,
- ,
- 巻
- 65
- 号
- 21
- 開始ページ
- 2999
- 終了ページ
- 3003
- 記述言語
- 英語
- 掲載種別
- DOI
- 10.1021/ac00069a011
A micro laser ablation system for the elemental analysis of a small area down to several microns in diameter was constructed, and its performance was investigated in combination with an inductively coupled plasma (ICP) mass spectrometry technique. A Q-switched, low-power Nd-YAG laser was used to ablate the samples. The products formed by laser ablation processes were trapped on a filter and then examined by scanning electron microscope. Entrainment and transport of ablated particulates were evaluated experimentally, and the detection limits for elements in an alloy sample were determined. © 1993, American Chemical Society. All rights reserved.
- リンク情報
- ID情報
-
- DOI : 10.1021/ac00069a011
- ISSN : 1520-6882
- ISSN : 0003-2700
- SCOPUS ID : 0027695858