Suitable hemolysis index for low-flow rotary blood pumps
JOURNAL OF ARTIFICIAL ORGANS
- ,
- ,
- ,
- ,
- 巻
- 24
- 号
- 2
- 開始ページ
- 120
- 終了ページ
- 125
- 記述言語
- 英語
- 掲載種別
- 研究論文(学術雑誌)
- DOI
- 10.1007/s10047-020-01218-4
- 出版者・発行元
- SPRINGER JAPAN KK
A suitable index is needed for hemolysis tests that use low-flow pumps, such as pediatric blood pumps or blood purification pumps. To create such an index, the present study investigates the change of plasma-free hemoglobin in the pump circuit with time and the change of the hemolysis rate with flow rate and impeller rotational speed. The results show that the hemolysis rate or the increase rate of the total free hemoglobin are suitable measures for hemolysis evaluation for low-flow pumps.