Frequency spectrum analysis of SJM cardiac prosthetic closing sound at mitral position.

Journal of the Japanese Society for Artificial Organs and Tissues
  • FUJITA Y.
  • ,
  • HAYASHI J.
  • ,
  • MORO H.
  • ,
  • SAITO A.
  • ,
  • UENO M.
  • ,
  • EGUCHI S.
  • ,
  • HORI J.
  • ,
  • SAITO Y.

Volume
19
Number
1
First page
382
Last page
385
Language
Japanese
Publishing type
DOI
10.11392/jsao1972.19.382
Publisher
JAPANESE SOCIETY FOR ARTIFICIAL ORGANS

The closing sounds of eleven St. Jude Medical cardiac prostheses at mitral position were analysed using the technique of the fast Fourier transform. The frequency-spectra contain a peak at about 1KHz, and decline the power to 5 to 7KHz. In two patients, 8.3 years and 4.5 years postoperatively, their frequency spectra of one leaflet showed a dip at 2.5KHz, respectively. The closing sound analysis is a useful methodd for the early detection of thrambosed SJM cardiac prosthesis at mitral position as well as at aortic or pulmonary position.