Improvement of precision in quantitative measurements of radioactivity with an imaging plate

APPLIED RADIATION AND ISOTOPES
  • T Suzuki
  • ,
  • C Mori
  • ,
  • H Miyahara
  • ,
  • A Uritani
  • ,
  • M Yoshida
  • ,
  • F Takahashi

49
9-11
開始ページ
1127
終了ページ
1130
記述言語
英語
掲載種別
DOI
10.1016/S0969-8043(97)10032-X
出版者・発行元
PERGAMON-ELSEVIER SCIENCE LTD

To obtain a high precision in radiation or radioactivity measurements with an imaging plate, the photo-stimulated luminescence intensity must be large. To get the highest intensity, we reread an imaging plate repeatedly and summed each intensity for the purpose of reducing the variation of the measured intensities. A great improvement of the precision in the measurement was achieved with the method of repeated readings when the time interval between each reading is more than about 1 h. (C) 1998 Elsevier Science Ltd. All rights reserved.