MISC

2006年8月

Low-level radiation measurement system with magnetically levitated electrode ionization chamber detector

IEEE TRANSACTIONS ON NUCLEAR SCIENCE
  • T. Kawaguchi
  • ,
  • K. Futagami
  • ,
  • M. Matoba
  • ,
  • G. Wakabayashi
  • ,
  • N. Ikeda
  • ,
  • Y. Uozumi
  • ,
  • N. Yamanaka
  • ,
  • M. Kaneko

53
4
開始ページ
2276
終了ページ
2280
記述言語
英語
掲載種別
DOI
10.1109/TNS.2006.878572
出版者・発行元
IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC

An automatic system was developed for the measurement of radiation using a magnetically levitated electrode ionization chamber. The system was used to measure background and K-40 gamma ray dose rates. The background dose rate measurement demonstrated high sensitivity and reliability of the system, and a low-level 4 K gamma ray dose rate was successfully measured: For the case of a 30-minute measurement, the detection limit of the system is 13 nGy/h. This measurement system will be useful as an area monitor to measure space dose rates.

リンク情報
DOI
https://doi.org/10.1109/TNS.2006.878572
CiNii Articles
http://ci.nii.ac.jp/naid/80018937698
Web of Science
https://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=JSTA_CEL&SrcApp=J_Gate_JST&DestLinkType=FullRecord&KeyUT=WOS:000240242900032&DestApp=WOS_CPL
ID情報
  • DOI : 10.1109/TNS.2006.878572
  • ISSN : 0018-9499
  • CiNii Articles ID : 80018937698
  • Web of Science ID : WOS:000240242900032

エクスポート
BibTeX RIS