2006年8月
Low-level radiation measurement system with magnetically levitated electrode ionization chamber detector
IEEE TRANSACTIONS ON NUCLEAR SCIENCE
- ,
- ,
- ,
- ,
- ,
- ,
- ,
- 巻
- 53
- 号
- 4
- 開始ページ
- 2276
- 終了ページ
- 2280
- 記述言語
- 英語
- 掲載種別
- DOI
- 10.1109/TNS.2006.878572
- 出版者・発行元
- IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
An automatic system was developed for the measurement of radiation using a magnetically levitated electrode ionization chamber. The system was used to measure background and K-40 gamma ray dose rates. The background dose rate measurement demonstrated high sensitivity and reliability of the system, and a low-level 4 K gamma ray dose rate was successfully measured: For the case of a 30-minute measurement, the detection limit of the system is 13 nGy/h. This measurement system will be useful as an area monitor to measure space dose rates.
- リンク情報
- ID情報
-
- DOI : 10.1109/TNS.2006.878572
- ISSN : 0018-9499
- CiNii Articles ID : 80018937698
- Web of Science ID : WOS:000240242900032