PINフォトダイオード線量計を使用したマルチスライスx線CT検査における胸部被曝線量の測定

日本放射線技術学会誌
  • 青山 隆彦
  • ,
  • 小山 修司
  • ,
  • 成田 憲彦
  • ,
  • 伊藤 茂樹
  • ,
  • 阿知波 正剛
  • ,
  • 野口 由美子

56
12
開始ページ
1437
終了ページ
1446
記述言語
日本語
掲載種別
DOI
10.6009/jjrt.KJ00003110933
出版者・発行元
公益社団法人 日本放射線技術学会

Organ and tissue doses delivered during chest CT examinations were measured as a function of scanning parameters including tube current, slice thickness, and pitch of the multi-slice CT scanner. Measurements were carried out by using a chest dosimetry phantom that incorporated lung and vertebra models, and by putting PIN photodiode dosimeters into the phantom at sites of the esophagus or stomach, lung, bone marrow, and skin. Absorbed doses at each site in the phantom measured for a constant scan area indicated that these values could be precisely calculated from the value obtained at any scanning parameter by using inverse proportionality of dose to pitch and the data of x-ray intensity as a function of tube current and slice thickness measured for the CT scanner used. Relative dose at each site was found to be constant regardless of the scanning parameters, indicating that the dose value at each site could be calculated from that measured at only one point in the phantom. Effective doses and effective dose equivalents were estimated from organ or tissue doses for chest CT examinations, and were found to be 1.5 to 3 times larger than the average values of effective dose equivalents appearing in the literature.

リンク情報
DOI
https://doi.org/10.6009/jjrt.KJ00003110933
CiNii Articles
http://ci.nii.ac.jp/naid/110003460176
CiNii Books
http://ci.nii.ac.jp/ncid/AN00197784
URL
http://search.jamas.or.jp/link/ui/2001139832
ID情報
  • DOI : 10.6009/jjrt.KJ00003110933
  • ISSN : 0369-4305
  • CiNii Articles ID : 110003460176
  • CiNii Books ID : AN00197784

エクスポート
BibTeX RIS