MICRODOSIMETRIC EVALUATION OF THE NEUTRON FIELD FOR BNCT AT KYOTO UNIVERSITY REACTOR BY USING THE PHITS CODE

RADIATION PROTECTION DOSIMETRY
  • H. Baba
  • ,
  • Y. Onizuka
  • ,
  • M. Nakao
  • ,
  • M. Fukahori
  • ,
  • T. Sato
  • ,
  • Y. Sakurai
  • ,
  • H. Tanaka
  • ,
  • S. Endo

143
2-4
開始ページ
528
終了ページ
532
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1093/rpd/ncq511
出版者・発行元
OXFORD UNIV PRESS

In this study, microdosimetric energy distributions of secondary charged particles from the (10)B(n,alpha)(7)Li reaction in boron-neutron capture therapy (BNCT) field were calculated using the Particle and Heavy Ion Transport code System (PHITS). The PHITS simulation was performed to reproduce the geometrical set-up of an experiment that measured the microdosimetric energy distributions at the Kyoto University Reactor where two types of tissue-equivalent proportional counters were used, one with A-150 wall alone and another with a 50-ppm-boron-loaded A-150 wall. It was found that the PHITS code is a useful tool for the simulation of the energy deposited in tissue in BNCT based on the comparisons with experimental results.