論文

査読有り
2017年9月

Dual Isotope SPECT Study With Epilepsy Patients Using Semiconductor SPECT System

CLINICAL NUCLEAR MEDICINE
  • Tohru Shiga
  • Atsuro Suzuki
  • Kotarou Sakurai
  • Tsugiko Kurita
  • Wataru Takeuchi
  • Takuya Toyonaga
  • Kenji Hirata
  • Keiji Kobashi
  • Chietsugu Katoh
  • Naoki Kubo
  • Nagara Tamaki
  • 全て表示

42
9
開始ページ
663
終了ページ
668
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1097/RLU.0000000000001732
出版者・発行元
LIPPINCOTT WILLIAMS & WILKINS

Purpose We developed a prototype CdTe SPECT system with 4-pixel matched collimator for brain study. This system provides high-energy-resolution (6.6%), high-sensitivity (220 cps/MBq/head), and high-spatial-resolution images. The aim of this study was to evaluate dual-isotope study of CBF and central benzodiazepine receptor (BZR) images using Tc-99m-ECD and I-123-IMZ with the new SPECT system in patients with epilepsy comparing with single-isotope study using the conventional scintillation gamma camera.
Methods This study included 13 patients with partial epilepsy. The BZR images were acquired at 3 hours after I-123-IMZ injection for 20 minutes. The images of IMZ were acquired with a conventional 3-head scintillation gamma camera. After BZR image acquisition with the conventional camera, Tc-99m-ECD was injected, and CBF and BZR images were acquired simultaneously 5 minutes after ECD injection with the new SPECT system. The CBF images were also acquired with the conventional camera on separate days. The findings were visually analyzed, and 3D-SSP maximum Z scores of lesions were compared between the 2 studies.
Results There were 47 abnormal lesions on BZR images and 60 abnormal lesions on CBF images in the single-isotope study with the conventional camera. Dual-isotope study with the new system showed concordant abnormal findings of 46 of 47 lesions on BZR and 54 of 60 lesions on CBF images with the single-isotope study with the conventional camera. There was high agreement between the 2 studies in both BZR and CBF findings (Cohen values = 0.96 for BZR and 0.78 for CBF). In semiquantitative analysis, maximum Z scores of dual-isotope study with the new system strongly correlated with those of single-isotope study with the conventional camera (BZR: r = 0.82, P < 0.05, CBF: r = 0.87, P < 0.05).
Conclusions Our new SPECT system permits dual-isotope study for pixel-by-pixel analysis of CBF and BZR information with the same pathophysiological condition in patients with epilepsy.

リンク情報
DOI
https://doi.org/10.1097/RLU.0000000000001732
Web of Science
https://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=JSTA_CEL&SrcApp=J_Gate_JST&DestLinkType=FullRecord&KeyUT=WOS:000407889900002&DestApp=WOS_CPL
ID情報
  • DOI : 10.1097/RLU.0000000000001732
  • ISSN : 0363-9762
  • eISSN : 1536-0229
  • Web of Science ID : WOS:000407889900002

エクスポート
BibTeX RIS