Dual Isotope SPECT Study With Epilepsy Patients Using Semiconductor SPECT System
CLINICAL NUCLEAR MEDICINE
- 巻
- 42
- 号
- 9
- 開始ページ
- 663
- 終了ページ
- 668
- 記述言語
- 英語
- 掲載種別
- 研究論文(学術雑誌)
- DOI
- 10.1097/RLU.0000000000001732
- 出版者・発行元
- LIPPINCOTT WILLIAMS & WILKINS
Purpose We developed a prototype CdTe SPECT system with 4-pixel matched collimator for brain study. This system provides high-energy-resolution (6.6%), high-sensitivity (220 cps/MBq/head), and high-spatial-resolution images. The aim of this study was to evaluate dual-isotope study of CBF and central benzodiazepine receptor (BZR) images using Tc-99m-ECD and I-123-IMZ with the new SPECT system in patients with epilepsy comparing with single-isotope study using the conventional scintillation gamma camera.
Methods This study included 13 patients with partial epilepsy. The BZR images were acquired at 3 hours after I-123-IMZ injection for 20 minutes. The images of IMZ were acquired with a conventional 3-head scintillation gamma camera. After BZR image acquisition with the conventional camera, Tc-99m-ECD was injected, and CBF and BZR images were acquired simultaneously 5 minutes after ECD injection with the new SPECT system. The CBF images were also acquired with the conventional camera on separate days. The findings were visually analyzed, and 3D-SSP maximum Z scores of lesions were compared between the 2 studies.
Results There were 47 abnormal lesions on BZR images and 60 abnormal lesions on CBF images in the single-isotope study with the conventional camera. Dual-isotope study with the new system showed concordant abnormal findings of 46 of 47 lesions on BZR and 54 of 60 lesions on CBF images with the single-isotope study with the conventional camera. There was high agreement between the 2 studies in both BZR and CBF findings (Cohen values = 0.96 for BZR and 0.78 for CBF). In semiquantitative analysis, maximum Z scores of dual-isotope study with the new system strongly correlated with those of single-isotope study with the conventional camera (BZR: r = 0.82, P < 0.05, CBF: r = 0.87, P < 0.05).
Conclusions Our new SPECT system permits dual-isotope study for pixel-by-pixel analysis of CBF and BZR information with the same pathophysiological condition in patients with epilepsy.
Methods This study included 13 patients with partial epilepsy. The BZR images were acquired at 3 hours after I-123-IMZ injection for 20 minutes. The images of IMZ were acquired with a conventional 3-head scintillation gamma camera. After BZR image acquisition with the conventional camera, Tc-99m-ECD was injected, and CBF and BZR images were acquired simultaneously 5 minutes after ECD injection with the new SPECT system. The CBF images were also acquired with the conventional camera on separate days. The findings were visually analyzed, and 3D-SSP maximum Z scores of lesions were compared between the 2 studies.
Results There were 47 abnormal lesions on BZR images and 60 abnormal lesions on CBF images in the single-isotope study with the conventional camera. Dual-isotope study with the new system showed concordant abnormal findings of 46 of 47 lesions on BZR and 54 of 60 lesions on CBF images with the single-isotope study with the conventional camera. There was high agreement between the 2 studies in both BZR and CBF findings (Cohen values = 0.96 for BZR and 0.78 for CBF). In semiquantitative analysis, maximum Z scores of dual-isotope study with the new system strongly correlated with those of single-isotope study with the conventional camera (BZR: r = 0.82, P < 0.05, CBF: r = 0.87, P < 0.05).
Conclusions Our new SPECT system permits dual-isotope study for pixel-by-pixel analysis of CBF and BZR information with the same pathophysiological condition in patients with epilepsy.