論文

査読有り
2006年1月

Peri-implant stress analysis in simulation models with or without trabecular bone structure

INTERNATIONAL JOURNAL OF PROSTHODONTICS
  • R Stegaroiu
  • ,
  • N Watanabe
  • ,
  • M Tanaka
  • ,
  • S Ejiri
  • ,
  • S Nomura
  • ,
  • O Miyakawa

19
1
開始ページ
40
終了ページ
42
記述言語
英語
掲載種別
研究論文(学術雑誌)
出版者・発行元
QUINTESSENCE PUBLISHING CO INC

Most 3-dimensional (3D) finite element analyses (FEAs) simplify the cancellous bone to a block and completely ignore its trabecular structure. Thus, a 3D FEA was performed to compare the peri-implant stress distribution of a model in which the trabecular structure was accurately simulated (precise model) with that of a model with a homogenous cancellous bone component (simplified model). In contrast to the simplified model, the distribution patterns and higher stresses in the precise model may explain the overall bone resorption at the implant-bone interface in load-related implant failures. Further studies using data from the jawbone and a more detailed implant simulation are planned.