Nov 4, 2013
6PM3-PMN-050 Optimization of electrode design for a QCM based odor sensor using thermal desorption spectroscopy
- ,
- ,
- ,
- Volume
- 2013
- Number
- 5
- First page
- 225
- Last page
- 226
- Language
- Japanese
- Publishing type
- Publisher
- The Japan Society of Mechanical Engineers
We developed a single microheater combined quartz crystal microbalance array for thermal desorption spectroscopy. In this study, new designs of the microheater were fabricated and frequency stability of fabricated sensor was compared between the new designs of the microheater. In a design of the microheater, the frequency stability (3 σ) calculated from the frequency change for 60 sec was 1.4 Hz when the sensor was heated to a temperature of 140 degrees Celsius. The design was effective to improve the frequency stability under heating. In addition, identification between methanol and ethanol was carried out by using the sensor.
- Link information
-
- CiNii Articles
- http://ci.nii.ac.jp/naid/110009961109
- CiNii Books
- http://ci.nii.ac.jp/ncid/AA12456946
- ID information
-
- CiNii Articles ID : 110009961109
- CiNii Books ID : AA12456946