Misc.

Nov 4, 2013

6PM3-PMN-050 Optimization of electrode design for a QCM based odor sensor using thermal desorption spectroscopy

  • Sakai Jumpei
  • ,
  • Iida Naoyuki
  • ,
  • Sohgawa Masayuki
  • ,
  • Abe Takashi

Volume
2013
Number
5
First page
225
Last page
226
Language
Japanese
Publishing type
Publisher
The Japan Society of Mechanical Engineers

We developed a single microheater combined quartz crystal microbalance array for thermal desorption spectroscopy. In this study, new designs of the microheater were fabricated and frequency stability of fabricated sensor was compared between the new designs of the microheater. In a design of the microheater, the frequency stability (3 σ) calculated from the frequency change for 60 sec was 1.4 Hz when the sensor was heated to a temperature of 140 degrees Celsius. The design was effective to improve the frequency stability under heating. In addition, identification between methanol and ethanol was carried out by using the sensor.

Link information
CiNii Articles
http://ci.nii.ac.jp/naid/110009961109
CiNii Books
http://ci.nii.ac.jp/ncid/AA12456946
ID information
  • CiNii Articles ID : 110009961109
  • CiNii Books ID : AA12456946

Export
BibTeX RIS