MISC

2013年11月4日

6PM3-PMN-050 昇温脱離式QCMにおいセンサの形状最適化(マイクロセンサとナノ計測,ポスターセッション)

マイクロ・ナノ工学シンポジウム
  • 酒井 盾平
  • ,
  • 飯田 尚之
  • ,
  • 寒川 雅之
  • ,
  • 安部 隆

2013
5
開始ページ
225
終了ページ
226
記述言語
日本語
掲載種別
出版者・発行元
一般社団法人日本機械学会

We developed a single microheater combined quartz crystal microbalance array for thermal desorption spectroscopy. In this study, new designs of the microheater were fabricated and frequency stability of fabricated sensor was compared between the new designs of the microheater. In a design of the microheater, the frequency stability (3 σ) calculated from the frequency change for 60 sec was 1.4 Hz when the sensor was heated to a temperature of 140 degrees Celsius. The design was effective to improve the frequency stability under heating. In addition, identification between methanol and ethanol was carried out by using the sensor.