Comparison of a digital flat-panel versus screen-film, photofluorography and storage-phosphor systems by detection of simulated lung adenocarcinorna lesions using hard copy images

BRITISH JOURNAL OF RADIOLOGY
  • K Ono
  • ,
  • T Yoshitake
  • ,
  • K Akahane
  • ,
  • Y Yamada
  • ,
  • T Maeda
  • ,
  • M Kai
  • ,
  • T Kusama

78
934
開始ページ
922
終了ページ
927
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1259/bjr/75926824
出版者・発行元
BRITISH INST RADIOLOGY

The purpose of this study was to compare hard copy images from a flat-panel detector digital radiography system with conventional radiography, photofluorographic radiography and storage phosphor radiography for the detection of simulated lung adenocarcinoma lesions and also for radiation dose. To test the diagnostic performance of these four systems, the authors used 15 types of lung adenocarcinoma phantom according to Noguchi's classification and an anthropomorphic chest phantom. The visual evaluation of tumour detectability by four radiologists and two general thoracic surgeons was examined with a five-level confidence scale. Lung doses were measured with glass dosemeters for the chest radiology systems under the conditions used by each hospital and centre. Our results indicated that flat-panel detector digital radiography and storage phosphor radiography are not necessarily superior to conventional radiography and photofluorographic radiography for detecting lung adenocarcinomas when only hard copy images are used, and this suggests a need to carefully optimize chest radiography.

リンク情報
DOI
https://doi.org/10.1259/bjr/75926824
PubMed
https://www.ncbi.nlm.nih.gov/pubmed/16177015
Web of Science
https://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=JSTA_CEL&SrcApp=J_Gate_JST&DestLinkType=FullRecord&KeyUT=WOS:000232716000009&DestApp=WOS_CPL
ID情報
  • DOI : 10.1259/bjr/75926824
  • ISSN : 0007-1285
  • PubMed ID : 16177015
  • Web of Science ID : WOS:000232716000009

エクスポート
BibTeX RIS