MISC

査読有り
2013年

Evaluation of angular dependence of neutron detection efficiencies of ZnS scintillator detectors

IEEE Nuclear Science Symposium Conference Record
  • T. Nakamura
  • ,
  • K. Toh
  • ,
  • T. Kawasaki
  • ,
  • K. Honda
  • ,
  • H. Suzuki
  • ,
  • M. Ebine
  • ,
  • A. Birumachi
  • ,
  • K. Sakasai
  • ,
  • K. Soyama
  • ,
  • M. Katagiri

記述言語
英語
掲載種別
DOI
10.1109/NSSMIC.2013.6829650
出版者・発行元
Institute of Electrical and Electronics Engineers Inc.

The detection efficiencies of ZnS neutron-sensitive scintillator detector were evaluated as functions of the neutron incidence angle on the scintillator and neutron wavelength using a pulsed neutron beam that generates neutron with wavelengths from 0.5 to 8 Å. ZnS/6LiF and ZnS/ 10B2O3 scintillator screens with thicknesses of 0.45 and 0.22 mm, respectively, were used for the evaluation. The detection efficiency of each scintillator was strongly influenced by the incidence angle and the neutron wavelength. A fitting model that incorporates both the neutron-absorption and light-escape probabilities reproduced the measured detection efficiencies very well. © 2013 IEEE.

リンク情報
DOI
https://doi.org/10.1109/NSSMIC.2013.6829650
Web of Science
https://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=JSTA_CEL&SrcApp=J_Gate_JST&DestLinkType=FullRecord&KeyUT=WOS:000347163503022&DestApp=WOS_CPL
ID情報
  • DOI : 10.1109/NSSMIC.2013.6829650
  • ISSN : 1095-7863
  • SCOPUS ID : 84904210365
  • Web of Science ID : WOS:000347163503022

エクスポート
BibTeX RIS