論文

査読有り
2009年6月

Characteristics comparison between a cyclotron-based neutron source and KUR-HWNIF for boron neutron capture therapy

NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS
  • H. Tanaka
  • Y. Sakurai
  • M. Suzuki
  • S. Masunaga
  • Y. Kinashi
  • G. Kashino
  • Y. Liu
  • T. Mitsumoto
  • S. Yajima
  • H. Tsutsui
  • A. Maruhashi
  • K. Ono
  • 全て表示

267
11
開始ページ
1970
終了ページ
1977
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1016/j.nimb.2009.03.095
出版者・発行元
ELSEVIER SCIENCE BV

At Kyoto University Research Reactor Institute (KURRI), 275 clinical trials of boron neutron capture therapy (BNCT) have been performed as of March 2006, and the effectiveness of BNCT has been revealed. In order to further develop BNCT, it is desirable to supply accelerator-based epithermal-neutron sources that can be installed near the hospital. We proposed the method of filtering and moderating fast neutrons, which are emitted from the reaction between a beryllium target and 30-MeV protons accelerated by a cyclotron accelerator, using an optimum moderator system composed of iron, lead, aluminum and calcium fluoride. At present, an epithermal-neutron source is under construction from June 2008. This system consists of a cyclotron accelerator, beam transport system, neutron-yielding target, filter, moderator and irradiation bed.
In this article, an overview of this system and the properties of the treatment neutron beam optimized by the MCNPX Monte Carlo neutron transport code are presented. The distribution of biological effect weighted dose in a head phantom compared with that of Kyoto University Research Reactor (KUR) is shown. It is confirmed that for the accelerator, the biological effect weighted dose for a deeply situated tumor in the phantom is 18% larger than that for KUR, when the limit dose of the normal brain is 10 Gyeq. The therapeutic time of the cyclotron-based neutron sources are nearly one-quarter of that of KUR. The cyclotron-based epithermal-neutron source is a promising alternative to reactor-based neutron sources for treatments by BNCT. (C) 2009 Elsevier B.V. All rights reserved.

リンク情報
DOI
https://doi.org/10.1016/j.nimb.2009.03.095
J-GLOBAL
https://jglobal.jst.go.jp/detail?JGLOBAL_ID=200902233082842965
Web of Science
https://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=JSTA_CEL&SrcApp=J_Gate_JST&DestLinkType=FullRecord&KeyUT=WOS:000267528400016&DestApp=WOS_CPL
ID情報
  • DOI : 10.1016/j.nimb.2009.03.095
  • ISSN : 0168-583X
  • J-Global ID : 200902233082842965
  • Web of Science ID : WOS:000267528400016

エクスポート
BibTeX RIS