Design and performance evaluation of a master controller for endovascular catheterization

INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY
  • Jin Guo
  • ,
  • Shuxiang Guo
  • ,
  • Takashi Tamiya
  • ,
  • Hideyuki Hirata
  • ,
  • Hidenori Ishihara

11
1
開始ページ
119
終了ページ
131
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1007/s11548-015-1211-4
出版者・発行元
SPRINGER HEIDELBERG

It is difficult to manipulate a flexible catheter to target a position within a patient's complicated and delicate vessels. However, few researchers focused on the controller designs with much consideration of the natural catheter manipulation skills obtained from manual catheterization. Also, the existing catheter motion measurement methods probably lead to the difficulties in designing the force feedback device. Additionally, the commercially available systems are too expensive which makes them cost prohibitive to most hospitals. This paper presents a simple and cost-effective master controller for endovascular catheterization that can allow the interventionalists to apply the conventional pull, push and twist of the catheter used in current practice.
A catheter-sensing unit (used to measure the motion of the catheter) and a force feedback unit (used to provide a sense of resistance force) are both presented. A camera was used to allow a contactless measurement avoiding additional friction, and the force feedback in the axial direction was provided by the magnetic force generated between the permanent magnets and the powered coil.
Performance evaluation of the controller was evaluated by first conducting comparison experiments to quantify the accuracy of the catheter-sensing unit, and then conducting several experiments to evaluate the force feedback unit. From the experimental results, the minimum and the maximum errors of translational displacement were 0.003 mm (0.01 %) and 0.425 mm (1.06 %), respectively. The average error was 0.113 mm (0.28 %). In terms of rotational angles, the minimum and the maximum errors were (0.33 %) and (6 %), respectively. The average error was (3.01 %). The force resolution was approximately 25 mN and a maximum current of 3A generated an approximately 1.5 N force.
Based on analysis of requirements and state-of-the-art computer-assisted and robot-assisted training systems for endovascular catheterization, a new master controller with force feedback interface was proposed to maintain the natural endovascular catheterization skills of the interventionalists.