2011年3月
Influence of index table accuracy on roundness calibration in the multi-step method using monte carlo simulation
MAPAN-JOURNAL OF METROLOGY SOCIETY OF INDIA
- 巻
- 26
- 号
- 1
- 開始ページ
- 37
- 終了ページ
- 46
- 記述言語
- 英語
- 掲載種別
- 研究論文(学術雑誌)
- DOI
- 10.1007/s12647-011-0004-7
- 出版者・発行元
- METROLOGY SOC INDIA
The precise and accurate roundness measurement is required year by year. The multi-step method is commonly used in the high-accuracy measurement for the roundness. The multi-step method is the self-calibration of the roundness capable of performing highly-advanced precise measurement. It depends on the reason that is able to separate measurement wave from the spindle rotation component and the form component of hemispherical master. But the mathematical models of the roundness measurement using that method are very complicated, describing numerous parameters and processes. So, recently, the Monte Carlo technique has attracted increasing attention as a very useful method for uncertainty estimation. This paper describes influences of the indexing angle accuracy and the noises generated by the roundness measuring machine, on the roundness measurements, both of which were analyzed by the Monte Carlo simulation. Furthermore, an example of hemispherical master is compared with the actual measurement data between when a reference guide plate is used utilizing the magnetic rotary encoder.
- リンク情報
- ID情報
-
- DOI : 10.1007/s12647-011-0004-7
- ISSN : 0970-3950
- Web of Science ID : WOS:000290544400004