Improved Design of the Exit of a Beam Shaping Assembly for an Accelerator-driven BNCT System in Nagoya University

JPS Conf. Proc.
  • Kazuya Sato
  • ,
  • Akira Uritani
  • ,
  • Kenichi Watanabe
  • ,
  • Sachiko Yoshihashi
  • ,
  • Atsushi Yamazaki
  • ,
  • Yoshiaki Kiyanagi
  • ,
  • Kazuki Tsuchida

22
開始ページ
011003-1 - 011002-7
終了ページ
記述言語
英語
掲載種別
速報,短報,研究ノート等(学術雑誌)
DOI
10.7566/JPSCP.22.011003

Development of an accelerator-driven compact neutron source for BNCT, Boron Neutron Capture Therapy, has started in Nagoya University. We will show the results of improved design around the neutron beam exit of the Beam Shaping Assembly. We proposed a use of an “Extension collimator” that can make it easier to set a patient in a comfortable position during the neutron irradiation. Additionally, we could greatly improve performance of the collimator by using a new nozzle structure called as “Ni implant nozzle”. We, consequently, achieved the neutron beam characteristics that almost satisfied all of the recommended values reported in IAEA-TECDOC-1223.