血圧計形状を目指した採血支援システムに関する研究

日本機械学会関東支部第22期総会・講演会予稿集,CD-ROM(GS-0203)
  • 齊藤 浩一

2016
22
開始ページ
"GS0203
終了ページ
1"-"GS0203-2"
記述言語
日本語
掲載種別
出版者・発行元
一般社団法人日本機械学会

The purpose of this study is development of a Blood Collection Assistance System (BCAS) for accident prevention. Previous studies have developed assistance units for vein detection and vein puncture, which are difficult operations. However, the large size (400×330×440mm) of the units tend to place a psychological burden on users and errors occurring in each unit are large. This study aimed to achieve downsizing the equipment, improvement of the needle control accuracy and integration of the control system. The puncture unit was downsized (195×170×210mm) with redesign. The error in vein detection was 2.6mm (max) which was larger than the target accuracy of 0.1 mm for safety puncture, depending on the position of near infrared illumination. The maximum errors of needle positioning, directing and puncturing were 0.09mm, 0.59° and 0.21 mm, respectively. As the standard deviation was small, it seems that optimization of the control system could reduce the error.