A Study of Contactless Energy Transmission for an Implantable Medical Device

J. Magn. Soc. Jpn
  • T. Komai
  • ,
  • F. Sato
  • ,
  • H. Matsuki
  • ,
  • T. Sato
  • ,
  • T. Sato

33
3
開始ページ
328
終了ページ
332
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.3379/msjmag.0903RH8035
出版者・発行元
The Magnetics Society of Japan

It is necessary for patients with implanted medical devices to have surgery in order to change devices whose battery cells have run out. We have suggested changing the battery into a secondary rechargeable one with contactless energy transmission, but it has been difficult to transmit electrical energy without increasing the eddy loss caused by the metal case of the device. We designed a rechargeable cardiac pacemaker with contactless electrical energy transmission and confirmed that it has a secondary power transmission efficiency of 26-40% at 0.4 W. However, an implantable cardioverter defibrillator (ICD) needs a larger volumetric battery than a cardiac pacemaker. It is necessary to transmit a large amount of power without increasing the eddy current loss, and it is important to estimate the eddy loss caused by the metal case.