論文

査読有り
2015年10月

High-resolution in-situ LDV monitoring system for measuring velocity distribution in blood vessel

OPTICS COMMUNICATIONS
  • Tomoaki Kyoden
  • ,
  • Shotaro Abe
  • ,
  • Hiroki Ishida
  • ,
  • Shunsuke Akiguchi
  • ,
  • Tsugunobu Andoh
  • ,
  • Yogo Takada
  • ,
  • Tsunenobu Teranishi
  • ,
  • Tadashi Hachiga

353
開始ページ
122
終了ページ
132
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1016/j.optcom.2015.04.075
出版者・発行元
ELSEVIER SCIENCE BV

We herein describe a cross-sectional multiple-point laser Doppler velocimetry (CS-MLDV) system for monitoring blood vessels that are sutured and connected during an operation. In order to observe the condition of a blood vessel during an operation, the previously developed linear MLDV (L-MLDV) system can realize velocity distribution imaging of the carotid artery in a living mouse by means of traverse laser light. We subsequently developed a CS-MLDV system, which can measure the instantaneous two-dimensional (2D) flow velocity, by upgrading the optical components and signal processing used in L-MLDV. The validity of the CS-MLDV results was verified through comparison with the results of a computational fluid dynamics (CFD) analysis. The results of the CFD analysis were similar to the experimental results obtained under the same flow field condition. Moreover, an instantaneous 2D velocity distribution can be obtained even for the case of flowing blood. Finally, we carried out in-vivo measurement in a mesenteric vessel of a mouse in order to demonstrate the potential of the CS-MLDV for use in surgery. (C) 2015 Elsevier B.V. All rights reserved.

リンク情報
DOI
https://doi.org/10.1016/j.optcom.2015.04.075
Web of Science
https://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=JSTA_CEL&SrcApp=J_Gate_JST&DestLinkType=FullRecord&KeyUT=WOS:000357552500020&DestApp=WOS_CPL
ID情報
  • DOI : 10.1016/j.optcom.2015.04.075
  • ISSN : 0030-4018
  • eISSN : 1873-0310
  • Web of Science ID : WOS:000357552500020

エクスポート
BibTeX RIS