Evaluation of the Calibration Method for Accurate Analysis of Dissolved Silica by Continuous Flow Analysis

Analytical Sciences
  • Chikako CHEONG
  • ,
  • Aya SAKAGUCHI
  • ,
  • Keisuke SUEKI
  • ,
  • Masaki OHATA

36
2
開始ページ
247
終了ページ
253
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.2116/analsci.19p291
出版者・発行元
Japan Society for Analytical Chemistry

For accurately determining nutrients in seawater by continuous flow analysis (CFA), the characteristic of the calibration curve was examined in detail. Under absorbance below 0.8, the calibration curve and the bracketing methods showed more accurate results that the bias fell below 0.5%. The analytical results of dissolved silica in seawater from the nutrient maximum layer of the Pacific Ocean obtained by the proposed methods showed good agreement with those obtained by an ion exclusion chromatography postcolumn absorption spectrophotometry (IEC-postcolumn) and an ion exclusion chromatography isotope dilution ICP mass spectrometry (IEC-ID-ICP-MS). From the results, the analysis of nutrients in seawater could be accurately carried out by CFA with an expanded uncertainty of below 1% using both the calibration curve and the bracketing methods with an appropriate absorbance range.