AN69ST membranes adsorb nafamostat mesylate and affect the management of anticoagulant therapy: a retrospective study

Journal of Intensive Care
  • Takahiro Hirayama
  • ,
  • Nobuyuki Nosaka
  • ,
  • Yasumasa Okawa
  • ,
  • Soichiro Ushio
  • ,
  • Yoshihisa Kitamura
  • ,
  • Toshiaki Sendo
  • ,
  • Toyomu Ugawa
  • ,
  • Atsunori Nakao

5
1
開始ページ
46
終了ページ
46
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1186/s40560-017-0244-x
出版者・発行元
Springer Science and Business Media LLC

BACKGROUND: In Japan, nafamostat mesylate (NM) is frequently used as an anticoagulant during continuous renal replacement therapy (CRRT). The dialyzer membrane AN69ST has been reported to adsorb NM and affect the management of anticoagulant therapy. However, the adsorbed amount has not yet been quantitatively assessed. Therefore, in this study, we evaluated the pre- and post-hemofilter prolongation of the activated clotting time (ACT) in patients with AN69ST and PS membranes. We also measured the adsorption of NM in three types of CRRT membranes using an experimental model. METHODS: In a study of patients who underwent CRRT using AN69ST or PS membranes in 2015 at the Advanced Emergency and Critical Care Center, Okayama University Hospital, pre- and post-hemofilter ACT measurements were extracted retrospectively, and the difference was calculated. In addition, AN69ST (sepXiris100), PS (HEMOFEEL SHG-1.0), and PMMA membranes (HEMOFEEL CH-1.0N) were used in an in vitro model of a dialysis circuit, and the concentrations of NM were measured in pre- and post-hemofilter membranes and filtrates. RESULTS: The ACT difference was significantly lower in the group using AN69ST membranes (p < 0.01). In the in vitro model (n = 4) with adsorption and filtration, the post-hemofilter and filtrate concentrations of NM in AN69ST membranes were significantly lower than those in the PS and PMMA membranes (p < 0.01). The NM adsorption clearance of the AN69ST membrane was significantly higher than that of the PS and PMMA membranes. CONCLUSIONS: The AN69ST membrane had higher NM adsorption than the PS and PMMA membranes. This may have resulted in the lower ACT difference in patients undergoing CRRT using the AN69ST membrane than in patients undergoing CRRT using PS or PMMA membranes.

リンク情報
DOI
https://doi.org/10.1186/s40560-017-0244-x
PubMed
https://www.ncbi.nlm.nih.gov/pubmed/28729905
PubMed Central
https://www.ncbi.nlm.nih.gov/pmc/articles/PMC5516335
URL
http://link.springer.com/content/pdf/10.1186/s40560-017-0244-x.pdf
ID情報
  • DOI : 10.1186/s40560-017-0244-x
  • eISSN : 2052-0492
  • PubMed ID : 28729905
  • PubMed Central 記事ID : PMC5516335

エクスポート
BibTeX RIS