2016年1月1日
A novel biodegradable external mesh stent improved long-term patency of vein grafts by inhibiting intimal–medial hyperplasia in an experimental canine model
General Thoracic and Cardiovascular Surgery
- ,
- ,
- ,
- ,
- ,
- ,
- ,
- ,
- ,
- 巻
- 64
- 号
- 1
- 開始ページ
- 1
- 終了ページ
- 9
- 記述言語
- 英語
- 掲載種別
- 研究論文(学術雑誌)
- DOI
- 10.1007/s11748-015-0591-2
- 出版者・発行元
- Springer-Verlag Tokyo
Objectives: Increased hemodynamic stress on vein grafts used in the arterial system is associated with vein graft disease. We determined whether a novel biodegradable external mesh stent could inhibit medial–intimal hyperplasia by suppressing hemodynamic stress on vein grafts and improve long-term patency. Methods: Twenty-four beagles underwent bilateral femoral interposition grafting using reversed femoral veins. Vein grafts were externally supported by a novel poly l-lactide-ε-caprolactone copolymer (P(LA/CL)) biodegradable mesh stent or a nonabsorbable mesh stent. Vein grafts with no reinforcement were used as controls. The grafts were harvested 6 and 12 months after implantation for morphometric and immunohistochemical assessment. Results: The endoluminal circumferential vein graft length was smaller in the P(LA/CL) and nonabsorbable groups (17.2 ± 2.9 and 19.0 ± 0.3 mm, respectively), than that in the control group (25.0 ± 2.6 mm, P <
0.01) at 12 months. The mean intimal–medial thickness was thinner in P(LA/CL) and nonabsorbable stent groups (0.18 ± 0.05 and 0.16 ± 0.05 mm, respectively), than that in the control group (0.30 ± 0.08 mm, P <
0.01). Differences in the intimal–medial thickness among the groups were associated with the magnitude of cellular proliferating activity. The graft patency ratio (100 %) was higher in the P(LA/CL) group than that in the nonabsorbable and control groups (72.2 and 63.6 %, respectively, P <
0.05). Conclusions: The biodegradable P(LA/CL) external mesh stent improved vein graft patency for 12 months and prevented vein graft dilatation and intimal hyperplasia associated with suppressed neointimal layer cellular proliferating activity.
0.01) at 12 months. The mean intimal–medial thickness was thinner in P(LA/CL) and nonabsorbable stent groups (0.18 ± 0.05 and 0.16 ± 0.05 mm, respectively), than that in the control group (0.30 ± 0.08 mm, P <
0.01). Differences in the intimal–medial thickness among the groups were associated with the magnitude of cellular proliferating activity. The graft patency ratio (100 %) was higher in the P(LA/CL) group than that in the nonabsorbable and control groups (72.2 and 63.6 %, respectively, P <
0.05). Conclusions: The biodegradable P(LA/CL) external mesh stent improved vein graft patency for 12 months and prevented vein graft dilatation and intimal hyperplasia associated with suppressed neointimal layer cellular proliferating activity.