2017年9月1日
Magnetization and demagnetization of magnetic dental attachments in a 3-T MRI system
Radiological Physics and Technology
- ,
- ,
- ,
- ,
- ,
- ,
- 巻
- 10
- 号
- 3
- 開始ページ
- 294
- 終了ページ
- 300
- 記述言語
- 英語
- 掲載種別
- 研究論文(学術雑誌)
- DOI
- 10.1007/s12194-017-0399-0
- 出版者・発行元
- Springer Tokyo
The purpose of this study was to evaluate the magnetization and demagnetization of magnetic dental attachments in a 3.0-T magnetic resonance imaging (MRI) scanner. A high-field-strength (3.0 T) MRI scanner (Achieva 3.0 T, Philips, Amsterdam, Netherlands) was used. Magnetic flux leakage was measured using a gaussmeter. To evaluate the magnetization and demagnetization of the magnetic assemblies and keepers caused by the moving in and out of the MRI scanner, the magnetic force of the attachments was measured before and after the table was moved in and out. Two settings were used wherein the magnetic assemblies and keepers were positioned on the table at angles of 0° or 90° with respect to the magnetic flux of the static magnetic field. The movement of the table was repeated 15 times. In addition, the retentive force of the magnetic dental attachments was measured after magnetic field exposure. The magnetic force of the magnetic attachments positioned at 0° decreased significantly after moving in and out of the MRI scanner (p <
0.05). In contrast, the magnetic force of the magnetic attachments positioned at 90° was stable after the movement of the table. The magnetic force of the keepers placed at both 0° and 90° was slightly increased after the movement of the table. At 0°, the retentive force of the magnet-keeper combinations decreased when the magnetic assembly was exposed to the strong magnetic flux of the MRI scanner. Therefore, the removal of all removable magnetic dentures is recommended before an MRI examination.
0.05). In contrast, the magnetic force of the magnetic attachments positioned at 90° was stable after the movement of the table. The magnetic force of the keepers placed at both 0° and 90° was slightly increased after the movement of the table. At 0°, the retentive force of the magnet-keeper combinations decreased when the magnetic assembly was exposed to the strong magnetic flux of the MRI scanner. Therefore, the removal of all removable magnetic dentures is recommended before an MRI examination.
- リンク情報
- ID情報
-
- DOI : 10.1007/s12194-017-0399-0
- ISSN : 1865-0341
- ISSN : 1865-0333
- PubMed ID : 28452002
- SCOPUS ID : 85018288323