Integrated pressure and temperature sensor with high immunity against external disturbance for flexible endoscope operation

JAPANESE JOURNAL OF APPLIED PHYSICS
  • Yusaku Maeda
  • ,
  • Kohei Maeda
  • ,
  • Hideki Kobara
  • ,
  • Hirohito Mori
  • ,
  • Hidekuni Takao

56
4
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.7567/JJAP.56.04CF09
出版者・発行元
IOP PUBLISHING LTD

In this study, an integrated pressure and temperature sensor device for a flexible endoscope with long-term stability in in vivo environments was developed and demonstrated. The sensor, which is embedded in the thin wall of the disposable endoscope hood, is intended for use in endoscopic surgery. The device surface is coated with a Cr layer to prevent photoelectronic generation induced by the strong light of the endoscope. The integrated temperature sensor allows compensation for the effect of the temperature drift on a pressure signal. The fabricated device pressure resolution is 0.4 mmHg; the corresponding pressure error is 3.2 mmHg. The packaged device was used in a surgical simulation in an animal experiment. Pressure and temperature monitoring was achieved even in a pH 1 acid solution. The device enables intraluminal pressure and temperature measurements of the stomach, which facilitate the maintenance of internal stomach conditions. The applicability of the sensor was successfully demonstrated in animal experiments. (C) 2017 The Japan Society of Applied Physics

リンク情報
DOI
https://doi.org/10.7567/JJAP.56.04CF09
Web of Science
https://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=JSTA_CEL&SrcApp=J_Gate_JST&DestLinkType=FullRecord&KeyUT=WOS:000414623100036&DestApp=WOS_CPL
URL
https://www.scopus.com/inward/record.uri?partnerID=HzOxMe3b&scp=85017102276&origin=inward
ID情報
  • DOI : 10.7567/JJAP.56.04CF09
  • ISSN : 0021-4922
  • eISSN : 1347-4065
  • SCOPUS ID : 85017102276
  • Web of Science ID : WOS:000414623100036

エクスポート
BibTeX RIS