Multi-laboratory evaluation of immunoaffinity LC-MS-based glucagon-like peptide-1 assay.
Bioanalysis
- 巻
- 15
- 号
- 4
- 開始ページ
- 207
- 終了ページ
- 218
- 記述言語
- 英語
- 掲載種別
- 研究論文(学術雑誌)
- DOI
- 10.4155/bio-2022-0240
Background: Although the fit-for-purpose approach has been proposed for biomarker assay validation, practical data should be compiled to facilitate the predetermination of acceptance criteria. Methods: Immunoaffinity LC-MS was used to analyze glucagon-like peptide-1 as a model biomarker in six laboratories. Calibration curve, carryover, parallelism, precision, relative accuracy and processed sample stability were evaluated, and their robustness among laboratories was assessed. The rat glucagon-like peptide-1 concentrations in four blinded samples were also compared. Results: The obtained results and determined concentrations in the blinded samples at all laboratories were similar, with a few exceptions, and robust, despite the difference in optimization techniques among laboratories. Conclusion: The results provide insights into the predefinition of the acceptance criteria of immunoaffinity LC-MS-based biomarker assays.