Measurement of nicotine in indoor air collected by alkaline-coated solid phase cartridge followed by GC-MS analysis

JOURNAL OF HEALTH SCIENCE
  • Ikue Saito
  • ,
  • Hiroshi Seto

53
1
開始ページ
53
終了ページ
59
記述言語
英語
掲載種別
研究論文(学術雑誌)
出版者・発行元
PHARMACEUTICAL SOC JAPAN

Nicotine has been used as a selective marker to evaluate environmental tobacco smoke (ETS). In this study, a simple and precise nicotine measurement method using a solid phase cartridge (Sep-Pak((R)) Plus PS-2) followed by GC-MS analysis was evaluated. Prior to use, the solid phase cartridge was coated with potassium hydroxide (KOH), and indoor air was then passed through it. After the air sampling, nicotine was eluted from the cartridge with 10 mu g/ml triethylamine/dichloromethane, and the eluent was analyzed by GC-MS. The nicotine recovery from the spiked cartridge was more than 80% for an air sampling time of 12 hr, and the method detection limit (MDL) for the air sampling with an air volume of 721 was 0.35 mu g/m(3). When the cartridge was used for indoor air measurement, the nicotine concentrations in the smoking room ranged from 12.9 mu g/m(3) to 86.6 mu g/m(3), which were proportional to the number of smoked cigarettes.