論文

2017年

Electrochemical impedance analysis on solid electrolyte oxygen sensor with gas and liquid reference electrodes for liquid LBE

SPECIAL ISSUE OF FOR THE FIFTH INTERNATIONAL SYMPOSIUM ON INNOVATIVE NUCLEAR ENERGY SYSTEMS
  • Pribadi Mumpuni Adhi
  • ,
  • Nariaki Okubo
  • ,
  • Atsushi Komatsu
  • ,
  • Masatoshi Kondo
  • ,
  • Minoru Takahashi

131
開始ページ
420
終了ページ
427
記述言語
英語
掲載種別
研究論文(国際会議プロシーディングス)
DOI
10.1016/j.egypro.2017.09.472
出版者・発行元
ELSEVIER SCIENCE BV

The ionic conductivity of solid electrolyte may insufficient, and the sensor output signal will deviate from the theoretical one in low temperature. The performance of oxygen sensor with Ag/air reference electrode (RE) and liquid Bi/Bi2O3 RE was tested in low-temperature LBE at 300 degrees - 450 degrees C and the charge transfer reactions impedance at the electrode-electrolyte interface was analyzed by electrochemical impedance analysis (EIS). After steady state condition, both of the sensors performed well and can be used at 300 degrees - 450 degrees C. Bi/Bi2O3 RE has lower impedance than Ag/air RE. Therefore, the response time of the oxygen sensor with Bi/Bi203 RE is faster than the oxygen sensor with Ag/air RE in the low-temperature region. (C) 2017 The Authors. Published by Elsevier Ltd.

リンク情報
DOI
https://doi.org/10.1016/j.egypro.2017.09.472
Web of Science
https://gateway.webofknowledge.com/gateway/Gateway.cgi?GWVersion=2&SrcAuth=JSTA_CEL&SrcApp=J_Gate_JST&DestLinkType=FullRecord&KeyUT=WOS:000425681200059&DestApp=WOS_CPL
ID情報
  • DOI : 10.1016/j.egypro.2017.09.472
  • ISSN : 1876-6102
  • Web of Science ID : WOS:000425681200059

エクスポート
BibTeX RIS