論文

査読有り
2014年5月

"Wedge effect" in stenosed coronary artery assessed using a PressureWire

HEART AND VESSELS
  • Kosuke Nakasuka
  • ,
  • Shigenori Ito

29
3
開始ページ
413
終了ページ
416
記述言語
英語
掲載種別
研究論文(学術雑誌)
DOI
10.1007/s00380-013-0385-3
出版者・発行元
SPRINGER

In this case report, we describe the recovery of abnormal coronary pressure waveform using a PressureWire Certus during percutaneous coronary intervention in a patient with severe stenosis in the proximal segment of the left anterior descending coronary artery. Since the diastolic pressure in the distal left anterior descending coronary artery was lower than that in the aorta, the pressure waveform before percutaneous coronary intervention represented the left ventricular pressure through a fistula due to a "wedge effect" in the stenosis as if the pulmonary artery wedge pressure determined by a Swan-Ganz catheter reflected the pressure in the distal portion (left atrium). We diagnosed this case coronary artery-left ventricular fistula based on the above findings. PressureWire Certus may be a valuable tool with which to estimate the hemodynamics in a patient with a coronary anomaly.